Constant O2 Value

Analyzer displays unchanging oxygen percentage

Possible causes:

  • Clogged or closed Flow Limiter
  • Obstruction in the inlet gas path
  • Reversed polarity of the O2 sensor - sensor cable connected "upside-down"
  • Simulated O2 value setup in the Analyzer menu [He/O2, SOLO]

Step-by-step troubleshooting:

  • Check the flow from the flow limiter - eg. using a glass full of water, look for a steady stream of bubbles.
  • [DNA] Remove and inspect inlet cap
  • [He/O2, SOLO] Check menu / configuration for simulated O2 value.
  • Check the sensor cable connection polarity - red wire matches the plus mark
  • Power cycle analyzer

When to contact service: If value remains constant after cleaning and no troubleshooting steps help. Get photos of the "internal values" screens before contacting support.