Constant O2 Value
Analyzer displays unchanging oxygen percentage
Possible causes:
- Clogged or closed Flow Limiter
- Obstruction in the inlet gas path
- Reversed polarity of the O2 sensor - sensor cable connected "upside-down"
- Simulated O2 value setup in the Analyzer menu [He/O2, SOLO]
Step-by-step troubleshooting:
- Check the flow from the flow limiter - eg. using a glass full of water, look for a steady stream of bubbles.
- [DNA] Remove and inspect inlet cap
- [He/O2, SOLO] Check menu / configuration for simulated O2 value.
- Check the sensor cable connection polarity - red wire matches the plus mark
- Power cycle analyzer
When to contact service: If value remains constant after cleaning and no troubleshooting steps help. Get photos of the "internal values" screens before contacting support.