Analyzer shows constant O₂ value

If your analyzer displays a constant O₂ value that does not change during measurement, follow the steps below.

Analyzer shows constant O₂ value

If your analyzer displays a constant O₂ value that does not change during measurement, follow the steps below.

Possible causes

  • Blocked or restricted gas flow

  • Closed or partially closed flow limiter

  • Depleted or unresponsive oxygen sensor

  • Electronics malfunction

What to try

  • Verify that gas is flowing properly through the analyzer.

  • Check the flow limiter and ensure it is open and correctly adjusted.

  • Confirm oxygen sensor voltage in ambient air (typically around 8–13 mV depending on sensor type).

  • Replace the oxygen sensor if it is end-of-life or shows abnormal voltage.

  • Restart the analyzer and repeat the measurement.

When to contact service

If the O₂ value remains constant despite confirmed gas flow and a functioning oxygen sensor, please contact an authorized Divesoft Service Point.

You can find your nearest authorized service location on our Service Centers page.