Skip to content
English
  • There are no suggestions because the search field is empty.

DNA shows unstable or decreasing oxygen value

If your DNA analyzer shows unstable oxygen readings or the oxygen percentage gradually decreases during Nitrox measurement, follow the steps below.

Possible causes

  • Excessive gas flow

  • Depleted or aged O₂ sensor

  • Outdated DivesoftApp


What to try

  • Reduce the gas flow as much as possible. DNA measures accurately even at very low flow.

  • Do not press the DNA tightly against the valve. Angle it slightly to allow some gas to escape around the analyzer.

  • Replace the oxygen sensor if it is older than 5 years or if the voltage in ambient air is below 6 mV.

  • Check in the App Store or Google Play whether a DivesoftApp update is available.

When to contact service

If the issue persists after trying the steps above, please contact your nearest authorized Divesoft Service Point and provide details about your measurement conditions.

You can find your nearest authorized service location on our Service Centers page.