ERROR 3 - Weak oxygen sensor

ERROR 3 on startup is caused by oxygen sensor voltage being lower than 5mV

Possible causes:

  • Oxygen sensor end of life
  • Residual hypoxic mix in the measuring path
  • Corrupted factory calibration

Step-by-step troubleshooting:

  • Check oxygen sensor date of manufacture, if is the sensor older than 24 months consider replacing it.
  • Try replacing the oxygen sensor with a new or known good one.
  • If your last measured mixture was strongly hypoxic (close to 0% O2) flush the Analyzer with air, if necesary turn it off and back on.
  • If your Analyzer HW version is 4.0 or earlier - or FW version is 2.29 then check the Internal Values - look for "O2 zero adjust" value, if plain zero, then a factory calibration is necessary - contact support for further instructions.

When to contact service: If ERROR 3 continues even with known good oxygen sensor, contact your closest Divesoft Service Point.