Unstable He value / Flow Limiter High Flow

Flow limiter incorrect flow impacts mostly He reading - unstable or not measuring at all. But O2 readings can be affected as well, expecially with pure oxygen.

Possible causes:

  • Damaged flow limiter
  • Wrong flow rate setting
  • Damaged or failing helium sensor assembly

Step-by-step troubleshooting:

  • Check flow rate with flowmeter (or a glass full of water, there should be steady stream of bubbles, not a geyser), in numbers the flow should be 0.2-0.5 L/min.
    • Simple flow limiter's flow can be adjusted by turning the nozzle
    • Proffesional flow limiter cannot be adjusted, only closed or fully opened to release residual pressure
  • Inspect flow limiter for damage or modifications
  • Ensure steady, continuous flow during measurement
  • If your helium value stays unstable even after disconnecting the tubing from Flow limiter to Analyzer, the Analyzer requires further helium assembly troubleshooting, search for "internal values" and / or re-calibration procedure.

When to contact service: If flow cannot be regulated properly, or helium value stays unstable even after all troubleshooting steps.