How to perform full O₂ and He calibration on He/O₂ analyzer

If measurement values appear incorrect or unstable, performing both O₂ and He calibration may resolve the issue. Follow the procedure below carefully.

When to perform full calibration

  • Measurement values appear incorrect

  • Helium readings seem unstable

  • Only O₂ calibration was previously performed

  • Analyzer behavior suggests calibration inconsistency

Important before starting

  • The analyzer must be connected to a tank with air (approximately 21% O₂) during the first O₂ calibration.

  • The analyzer must also remain connected to air during He calibration.

  • Perform calibration in a stable environment with steady gas flow.

Calibration procedure

  1. Switch off the analyzer.

  2. Press and hold the power button.

  3. Wait until the version screen appears.

  4. Press and release the Mode button.

  5. Release the power button. The analyzer will start in Service Mode.

  6. Perform standard O₂ calibration.

  7. Navigate to Menu.

  8. Select the folder “Probe calibration”.

  9. Perform the calibration procedure.

  10. When the analyzer displays “Stable”, confirm by pressing OK. This completes the He calibration.

  11. Switch off the analyzer to save the calibration values.

  12. Switch the analyzer back on.

  13. Perform O₂ calibration again.

Expected result

After completing both O₂ and He calibration, the analyzer should operate correctly and provide stable measurement values.

When to contact service

If measurement issues persist after completing both O₂ and He calibration according to the procedure above, please contact an authorized Divesoft Service Point for further diagnostics.

You can find your nearest authorized service location on our Service Centers page.