Analyzer shows O2 overload error

If your analyzer displays an O₂ overload error while measuring pure oxygen, follow the steps below.

Possible causes

  • Non-genuine or incompatible oxygen sensor

  • Worn or end-of-life oxygen sensor

  • Incorrect electronic gain setting

What to try

  • Check the O₂ sensor voltage while measuring pure oxygen — for example, on the calibration screen. A healthy sensor should read within the expected voltage range specified in your device manual.

  • If the overload error appeared after replacing the oxygen sensor and your device is a He/O₂ hardware revision 3.x or earlier, a hardware gain adjustment is likely required. Do not attempt to adjust the gain yourself — contact an authorized Divesoft Service Point.

When to contact service

If the overload error persists after replacing the oxygen sensor, contact an authorized Divesoft Service Point. Devices on He/O₂ hardware revision 3.x or earlier require a hardware-level gain adjustment that must be performed by a qualified technician.

Find your nearest authorized service location on the Divesoft Service Centers page.