Analyzer shows O2 overload error

If your analyzer displays an O₂ overload error while measuring pure oxygen, follow the steps below.

Possible causes

  • Non-original or incompatible oxygen sensor

  • Aged or end-of-life oxygen sensor

  • Incorrect electronic gain setting

What to try

  • Check the actual O₂ sensor voltage while measuring pure oxygen (for example in the calibration screen).

  • If the overload error appeared after replacing the oxygen sensor and your device is a He/O₂ hardware revision 3.x or earlier, a hardware gain adjustment may be required.

When to contact service

If the overload error persists after replacing the oxygen sensor, or if your device is a He/O₂ hardware revision 3.x or earlier, please contact an authorized Divesoft Service Point.

You can find your nearest authorized service location on our Service Centers page.