Firmware Update - FLIP Timeout Error

Resolving FLIP timeout errors during He/O2 or SOLO Analyzer update

Possible causes:

  • Not holding the ON/OFF button during the entire firmware update
  • FLIP driver issues on the computer
  • USB cable or USB hub issues
  • Incorrect update procedure
  • Incorrect MCU type selection in FLIP (does not match the hardware revision)

Step-by-step troubleshooting:

  • Hold the ON/OFF button on the Analyzer throughout the entire firmware update process in FLIP
  • Double-check the MCU type selection in FLIP
    • Select AT89C51ED2 for hardware rev. 1.0 to 2.3
    • Select AT89C51RE2 for hardware rev. 2.4 and above
  • Install or reinstall the FLIP drivers (available from the Atmel/Microchip website)
  • Use a shorter USB cable (under 1.5 m / 5 ft) and connect directly to the computer — avoid USB hubs
  • Restart computer and analyzer
  • Follow the official firmware update guide for He/O2 or SOLO.

When to contact service: If the FLIP timeout error persists across multiple computers, contact our support team for further assistance.